BA02183C
SS500XP/SS2000XP Gas Analyzer
Hauser
B
–
9
11.
Dismount the cell from the bracket by removing the four securing
screws (two on top, two on the bottom) using a 9/64 in. Allen
wrench.
12.
Place the measurement cell on a clean, flat surface with the pressure
sensor facing up. Refer to Figure B–4.
13.
Using a 9/16 in. wrench, secure the flange while using a 7/8 in.
wrench to remove the old pressure sensor. Refer to Figure B–5.
Orient the measurement cell to avoid any debris from entering the
cell.
Figure B–4
Removed measurement cell with pressure
sensor face up
Figure B–5
Removing the old pressure
sensor