3 Installation
Proline Prosonic Flow 90
14
Hauser
3.2
Installation conditions
3.2.1
Installation dimensions
The dimensions and the fitting lengths of the sensors and the transmitter are on Page 104 ff.
3.2.2
Installation location
Correct measuring is possible only if the pipe is full.
Avoid
the following installation locations:
• Do not install at the highest point in the run. Risk of air accumulating.
• Do not install directly upstream from an open pipe outlet in a down pipe.
A0001103
Fig. 5:
Installation location
Down pipes
Notwithstanding the above, the installation proposal below permits installation in an open down
pipe. Pipe restrictions or the use of an orifice plate with a smaller cross-section than the nominal
diameter prevent the pipe from running empty while measurement is in progress.
A0001104
Fig. 6:
Installation in a down pipe
1 = Supply tank; 2 = Measuring sensors; 3 = Orifice plate, pipe restriction; 4 = Valve, 5 = Filling tank
1
2
3
4
5
Содержание Prosonic Flow 90
Страница 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Страница 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Страница 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Страница 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Страница 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Страница 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Страница 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Страница 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Страница 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...