3 Installation
Proline Prosonic Flow 90
30
Hauser
5.
Mark the sensor distance (a) from the middle line starting from the first drillhole.
6.
Project the middle line to the back of the pipe and draw it on.
A0001126
Fig. 26:
Installing the measuring sensors, steps 5 and 6
7.
Mark the drillhole on the middle line on the back of the pipe.
8.
Cut out the second drillhole and prepare the holes for welding of the sensor holder
(deburr, clean, etc.).
A0001127
Fig. 27:
Installing the measuring sensors, steps 7 and 8
9.
Insert the sensor holders into the two drill holes. To adjust the weld-in depth, both sensor
holders can be fixed with the special tool for insertion depth regulation (optional) and then
aligned using the tie rod. The sensor holder must be flush with the inner side of the pipe.
Now pinpoint both sensor holders.
!
Note!
To align the tie rod, two bearing shells must be screwed onto the sensor holders.
A000128
Fig. 28:
Installing the measuring sensors, step 9
a
Содержание Prosonic Flow 90
Страница 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Страница 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Страница 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Страница 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Страница 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Страница 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Страница 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Страница 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Страница 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...