Proline Prosonic Flow 90
9 Trouble-shooting
Hauser
83
No. # 6xx
→
Simulation operation active
S
!
POS. ZERO-RET.
# 601
Positive zero return active.
"
Caution!
This is the highest priority notice
message.
Switch off positive zero return
S
!
SIM. CURR. OUT. n
# 611...614
Simulation current output active
Switch off simulation
S
!
SIM. FREQ. OUT. n
# 621...624
Simulation frequency output active
Switch off simulation
S
!
SIM. PULSE n
# 631...634
Simulation pulse output active
Switch off simulation
S
!
SIM. STAT. OUT n
# 641...644
Simulation status output active
Switch off simulation
S
!
SIM.STATUS IN n
# 671...674
Simulation status input active
Switch off simulation
S
!
SIM. FAILSAFE
# 691
Simulation of response to error (out-
puts) active
Switch off simulation
S
!
SIM. MEASURAND
# 692
Simulation of volume flow active
Switch off simulation
S
!
DEV. TEST ACT.
# 698
The measuring device is being
checked on site via the test and simu-
lation device.
–
No. # 7xx
→
Calibration or action errors
S
!
ADJ.ZERO FAIL.
# 731
Static zero point adjustment is not
possible or was cancelled.
Check that the flow velocity
is = 0 m/s.
Type
Error message / No.
Cause
Remedy / spare part
Содержание Prosonic Flow 90
Страница 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Страница 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Страница 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Страница 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Страница 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Страница 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Страница 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Страница 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Страница 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...