3 Installation
Proline Prosonic Flow 90
32
Hauser
3.3.9
Installing the wall-mount housing,
There are various ways of installing the wall-mount transmitter housing:
• Direct wall mounting
• Panel mounting (with separate mounting kit, accessories
• Pipe mounting (with separate mounting kit, accessories
"
Caution!
• At the installation location, make sure that ambient temperature does not exceed the permissible
range (–20 °...+60 °C). Install the device at a shady location. Avoid direct sunlight.
• Always install the wall-mount housing in such a way that the cable entries are pointing down.
Direct wall mounting
1.
Drill the holes as illustrated in Fig. 30.
2.
Remove the cover of the connection compartment (a).
3.
Push the two securing screws (b) through the appropriate bores (c) in the housing.
– Securing screws (M6): max. Ø 6.5 mm
– Screw head: max. Ø 10.5 mm
4.
Secure the transmitter housing to the wall as indicated.
5.
Screw the cover of the connection compartment (a) firmly onto the housing.
A0001130
Fig. 30:
Mounted directly on the wall
90
35
a
b
192
81.5
c
c
Содержание Prosonic Flow 90
Страница 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Страница 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Страница 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Страница 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Страница 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Страница 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Страница 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Страница 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Страница 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...