6 Commissioning
Proline Prosonic Flow 90
68
Hauser
Procedure (determining data for sensor installation)
You can use the following tables to select and configure, in the correct order, the functions required
to install the sensor:
• “clamp-on” sensor installation
• “Insertion” sensor installation
!
Note!
You must enter a valid release code before device parameters can be changed or activated.
The code (factory setting = 90) is entered by means of the corresponding matrix cell.
“Clamp-on” sensor installation
Procedure
Selection - Input - display
Local display (Quick Setup)
▼
ToF Tool - Fieldtool Package
▼
▼
→
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR PARAM.
→
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR PARAM.
Type of measurement
MEASUREMENT
MEASUREMENT
Sensor type
SENSOR TYPE
SENSOR TYPE
Sensor configuration
SENSOR CONFIGURATION
SENSOR CONFIGURATION
▼
→
BASIC FUNCTION
→
PROCESS PARAM.
→
PIPE DATA
→
BASIC FUNCTION
→
PROCESS PARAM.
→
PIPE DATA
Standard pipe selection
PIPE STANDARD
PIPE STANDARD
Nominal diameter pipe
NOMINAL DIAMETER
NOMINAL DIAMETER
Pipe material
PIPE MATERIAL
PIPE MATERIAL
Pipe sound velocity
SOUND VELOCITY PIPE
SOUND VELOCITY PIPE
Pipe circumference
CIRCUMFERENCE
CIRCUMFERENCE
Pipe diameter
PIPE DIAMETER
PIPE DIAMETER
Wall thickness
WALL THICKNESS
WALL THICKNESS
Liner material
LINER MATERIAL
LINER MATERIAL
Liner sound velocity
SOUND VELOCITY LINER
SOUND VELOCITY LINER
Liner thickness
LINER THICKNESS
LINER THICKNESS
▼
→
BASIC FUNCTION
→
PROCESS PARAM.
→
LIQUID DATA
→
BASIC FUNCTION
→
PROCESS PARAM.
→
LIQUID DATA
Liquid in the pipe
LIQUID
LIQUID
Liquid temperature
TEMPERATURE
TEMPERATURE
Liquid sound velocity
SOUND VELOCITY LIQUID
SOUND VELOCITY LIQUID
▼
→
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR PARAM.
→
BASIC FUNCTION
→
SENSOR DATA
→
SENSOR PARAM.
Didsplay sensorposition
(for sensor installation)
POSITION SENSOR
POSITION SENSOR
Содержание Prosonic Flow 90
Страница 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Страница 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Страница 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Страница 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Страница 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Страница 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Страница 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Страница 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Страница 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...