Mounting
ProlineйPromagйDй400йHART
28
Hauser
9. Removeйtheйelectronicsйmoduleйfromйtheйmainйelectronicsйmodule.
10. Loosenйtheйfixingйscrewsйofйtheйtransmitterйhousingйсwhenйreassembling,йpay
attentionйtoйtheйtighteningйtorqueйс ййй28)).
11. Liftйtheйtransmitterйhousing.
12. Turnйtheйhousingйtoйtheйdesiredйpositionйinйincrementsйofй90°.
Reassembling the transmitter housing
WARNING
Excessive tightening torque applied to the fixing screws!
Damageйtoйtheйtransmitter.
‣
Whenйreassembling,йtightenйtheйfixingйscrewsйasйperйtheйtighteningйtorque:
Step
Fixing screw
Tightening torques for housing made of:
Aluminum
Plastic
1
Housingйcover
2.5йNmйс1.8йlbfйft)
1йNmйс0.7йlbfйft)
5
Smartйsensorйelectronicsйmodule
0.6йNmйс0.4йlbfйft)
7
Mainйelectronicsйmodule
1.5йNmйс1.1йlbfйft)
10
Transmitterйhousing
5.5йNmйс4.1йlbfйft)
NOTICE
Plug of the smart sensor electronics module connected incorrectly!
Noйmeasuringйsignalйisйoutput.
‣
Plugйinйtheйplugйofйtheйsmartйsensorйelectronicsйmoduleйasйperйtheйcoding.
ййA0021585
NOTICE
Incorrect routing of the connecting cables between the sensor and transmitter in the
transmitter housing!
Thisйcanйinterfereйwithйtheйmeasuringйsignal.
‣
Routeйtheйconnectingйcablesйdirectlyйatйtheйlevelйofйtheйplugs.
‣
Reverseйtheйprocedureйtoйreassembleйtheйmeasuringйdevice.
6.2.6
Turning the display module
Theйdisplayйmoduleйcanйbeйturnedйtoйoptimizeйdisplayйreadabilityйandйoperability.
Содержание Proline Promag D 400 HART