Commissioning
ProlineйPromagйDй400йHART
90
Hauser
10.5.2 Carrying out a sensor adjustment
Theй
Sensor adjustment
йsubmenuйcontainsйparametersйthatйpertainйtoйtheйfunctionalityйof
theйsensor.
Navigation
"Setup"йmenuй йAdvancedйsetupй йSensorйadjustment
Structure of the submenu
Sensor adjustment
Installationйdirection
Parameter overview with brief description
Parameter
Description
Selection
Factory setting
Installationйdirection
Setйsignйofйflowйdirectionйtoйmatchйthe
directionйofйtheйarrowйonйtheйsensor.
• Flowйinйarrowйdirection
• Flowйagainstйarrowйdirection
Flowйinйarrowйdirection
10.5.3 Configuring the totalizer
Inйtheй
"Totalizer 1 to 3" submenu
йtheйindividualйtotalizersйcanйbeйconfigured.
Navigation
"Setup"йmenuй йAdvancedйsetupй йTotalizerй1йtoй3
Totalizer 1 to 3
Assignйprocessйvariable
Unitйtotalizer
Totalizerйoperationйmode
Failureйmode
Parameter overview with brief description
Parameter
Description
Selection
Factory setting
Assignйprocessйvariable
Selectйprocessйvariableйforйtotalizer.
• Off
• Volumeйflow
• Massйflow
Volumeйflow
Unitйtotalizer
Selectйprocessйvariableйtotalizerйunit.
Unitйchooseйlist
l
Totalizerйoperationйmode
Selectйtotalizerйcalculationйmode.
• Netйflowйtotal
• Forwardйflowйtotal
• Reverseйflowйtotal
Netйflowйtotal
Failureйmode
Defineйtotalizerйbehaviorйinйalarmйcondition. • Stop
• Actualйvalue
• Lastйvalidйvalue
Stop
Содержание Proline Promag D 400 HART