Commissioning
ProlineйPromagйDй400йHART
80
Hauser
Parameter
Description
Selection / User entry
Factory setting
Assignйlimit
Selectйprocessйvariableйforйlimitйfunction.
• Off
• Volumeйflow
• Massйflow
• Flowйvelocity
• Electronicйtemperature
• Totalizerй1
• Totalizerй2
• Totalizerй3
Volumeйflow
Assignйflowйdirectionйcheck
Selectйprocessйvariableйforйflowйdirection
monitoring.
• Off
• Volumeйflow
• Massйflow
Volumeйflow
Assignйstatus
Selectйdeviceйstatusйforйswitchйoutput.
• Emptyйpipeйdetection
• Lowйflowйcutйoff
Emptyйpipeйdetection
Massйflowйunit
Selectйmassйflowйunit.
Result
Theйselectedйunitйappliesйfor:
• Output
• Lowйflowйcutйoff
• Simulationйprocessйvariable
Unitйchooseйlist
Country-specific:
• kg/h
• lb/min
Volumeйflowйunit
Selectйvolumeйflowйunit.
Result
Theйselectedйunitйappliesйfor:
• Output
• Lowйflowйcutйoff
• Simulationйprocessйvariable
Unitйchooseйlist
Country-specific:
• l/h
• gal/minйсus)
Unitйtotalizer
Selectйprocessйvariableйtotalizerйunit.
Unitйchooseйlist
l
Switch-onйvalue
Enterйmeasuredйvalueйforйtheйswitch-on
point.
Signedйfloating-pointйnumber
0йl/h
Switch-offйvalue
Enterйmeasuredйvalueйforйtheйswitch-off
point.
Signedйfloating-pointйnumber
0йl/h
Switch-onйdelay
Defineйdelayйforйtheйswitch-onйofйstatus
output.
0.0йtoй100.0йs
0.0йs
Switch-offйdelay
Defineйdelayйforйtheйswitch-offйofйstatus
output.
0.0йtoй100.0йs
0.0йs
Failureйmode
Defineйoutputйbehaviorйinйalarmйcondition.
• Actualйstatus
• Open
• Closed
Open
Invertйoutputйsignal
Invertйtheйoutputйsignal.
• No
• Yes
No
10.4.5 Configuring the local display
Theй
Display
йwizardйguidesйyouйsystematicallyйthroughйallйtheйparametersйthatйcan
configuredйforйconfiguringйtheйlocalйdisplay.
Navigation
"Setup"йmenuй йDisplay
Содержание Proline Promag D 400 HART