ProlineйPromagйDй400йHART
Technicalйdata
Hauser
131
Currentйconsumption
Transmitter
Order code for "Power supply"
Maximum
Current consumption
Maximum
switch-on current
Optionй
L
:йACй100йtoй240йV
145йmA
25йAйс<й5йms)
Optionй
L
:йAC/DCй24йV
350йmA
27йAйс<й5йms)
Powerйsupplyйfailure
• Totalizersйstopйatйtheйlastйvalueйmeasured.
• Configurationйisйretainedйinйtheйplug-inйmemoryйсHistoROMйDAT).
• Errorйmessagesйсincl.йtotalйoperatedйhours)йareйstored.
Electricalйconnection
Potentialйequalization
Terminals
Transmitter
• Supplyйvoltageйcable:йplug-inйspringйterminalsйforйwireйcross-sections
0.5йtoй2.5йmm
2
йс20йtoй14йAWG)
• Signalйcable:йplug-inйspringйterminalsйforйwireйcross-sections
0.5йtoй2.5йmm
2
йс20йtoй14йAWG)
• Electrodeйcable:йspringйterminalsйforйwireйcross-sectionsй0.5йtoй2.5йmm
2
йс20йtoй14йAWG)
• Coilйcurrentйcable:йspringйterminalsйforйwireйcross-sections
0.5йtoй2.5йmm
2
йс20йtoй14йAWG)
Sensor connection housing
Springйterminalsйforйwireйcross-sections0.5йtoй2.5йmm
2
йс20йtoй14йAWG)
Cableйentries
Cable entry thread
• M20йxй1.5
• Viaйadapter:
– NPTй½"
– Gй½"
Cable gland
• Forйstandardйcable:йM20й×й1.5йwithйcableй 6йtoй12йmmйс0.24йtoй0.47йin)
• Forйreinforcedйcable:йM20й×й1.5йwithйcableй 9.5йtoй16йmmйс0.37йtoй0.63йin)
Ifйmetalйcableйentriesйareйused,йuseйaйgroundingйplate.
Cableйspecification
16.6 Performance characteristics
Referenceйoperating
conditions
In accordance with DIN EN 29104
• Fluidйtemperature:й+28±2й°Cйс+82±4й°F)
• Ambientйtemperatureйrange:й+22±2й°Cйс+72±4й°F)
• Warm-upйperiod:30йmin
Installation
• Inletйrunй>й10й×йDN
• Outletйrunй>й5й×йDN
• Sensorйandйtransmitterйgrounded.
• Theйsensorйisйcenteredйinйtheйpipe.