Commissioning
ProlineйPromagйDй400йHART
84
Hauser
Structure of the wizard
Volume flow
Assign freq.
Mass flow
Off
Volume flow
Assign pulse
Mass flow
Off
Volume flow
Assign curr.
Mass flow
Off
Mode output 3
Reverse flow*
Forward flow
Forward/Reverse
End of wizard
Display damping
Damping out. 2
Damping out. 1
Mode output 2
Mode output 1
Rev. flow comp.
ййA0023200-EN
30
"Output conditioning" wizard in the "Setup" menu
Reverse flow* = option only for pulse and frequency output
Parameter overview with brief description
Parameter
Description
User entry / Selection
Factory setting
Displayйdamping
Setйdisplayйreactionйtimeйtoйfluctuationsйin
theйmeasuredйvalue.
0.0йtoй999.9йs
0.0йs
Dampingйoutputй1
Setйreactionйtimeйforйoutputйsignalйto
fluctuationsйinйtheйmeasuredйvalue.
0йtoй999.9йs
1йs
Dampingйoutputй1
Setйreactionйtimeйforйoutputйsignalйto
fluctuationsйinйtheйmeasuredйvalue.
0йtoй999.9йs
1йs
Assignйcurrentйoutput
Selectйprocessйvariableйforйcurrentйoutput.
• Off
• Volumeйflow
• Massйflow
Volumeйflow
Assignйpulseйoutput
Selectйprocessйvariableйforйpulseйoutput.
• Off
• Volumeйflow
• Massйflow
Off