Interruption of the calibration operation
The calibration operation can be interrupted
at any time by pressing the MEAS key.
If the interruption is confirmed by “yes, abort
calibration”, the system returns to measuring
mode.
If “no” is chosen instead, the calibration
operation is continued.
Note:
If a retractable assembly is
connected, it has to be moved to
measuring position via the menu
“Basic operation”.
If there is a power failure in the course of the
calibration operation, the calibration operation
can be repeated or terminated when the
power supply has been re-established.
If “End calibration” is chosen, the same fields
appear as is the case when the calibration
operation has been completed.
General calibration instructions
Caution:
• Clean electrode before calibration.
• If symmetrical high-impedance
electrode connection is chosen,
potential matching line has to be
in connection with the buffer
solution too, e.g. by means of
a wire.
• If automatic temperature
compensation (ATC) is selected,
the relevant temperature sensors
must be immersed in the buffer
solution too.
Note:
• The fields marked with
a black triangle only appear
when they are required.
• Confirmation of input and
switching to the next field is
achieved by pressing the
key.
• When you see these symbols
in the operating table, immerse
one or both of the electrodes
(as shown on the display in each
individual case) in buffer
solution 1 or 2. Start calibration
with the
key.
• The only way to conclude the
calibration operation is with
“End of calib
.
2
E
↑
↑
pm152e09.chp
mycom CPM 152
Calibration