7.5.3
Buffer selection (pH)
7.5.4
Calibration parameters (Redox)
7.5.5
General settings (Redox)
Function
Selection
Default
Buffer type
E+H,
Ingold; Merck / Riedel;
DIN; special buffer type;
Japan; Fisher; Beckmann;
Cole-Parmer; OMEGA
E+H
If one of the standard buffers is chosen (only relevant for calibration with preset buffer)
Selection of buffer 1
Different pH values of the selected
buffer type
6.98
Selection of buffer 2
4.01
If a special buffer type is chosen
Number of implementable buffers
2 ... 3
3
– For buffer 1, 2 and if relevant 3 –
Number of reference points
2 ... 10
10
Buffer table
Numeric entry of max. 10
value pairs (temperature / pH)
for each buffer
(Intermediate values are
determined on a linear basis)
Selection of buffer 1
Special buffer 1, 2, 3
1
Selection of buffer 2
2
Function
Selection
Default
Calibration type
with Redox absolute (mV):
Data entry absolute
Calibration absolute
with Redox relative (%)
Data entry absolute
Data entry relative
Calibration absolute
Calibration relative
50 % end point
Calibration
absolute
Calibration
relative
Calibrate electrodes
(only with 2-circuit instrument,
except numerical calibration)
separately
together
separately
Function
Selection
Default
Hold during calibration
yes / no
yes
Alarm offset
1 ... 1500 mV
100 mV
pm152e07.chp
mycom CPM 152
Instrument configuration
Hauser
43