9.3.3
Calibration absolute (%)
Display
Description
The redox value measured with the old
calibration data is displayed. Wait until the
measurement has stabilised. Press the E key
to store the measurement.
Enter the absolute redox value of the buffer
in mV.
Calibration information. If there is a
calibration fault you will be informed by a
display.
Press the E key.
Final menu. You can:
•
store the calibration
•
repeat the calibration
•
discard the calibration and
return to measuring mode
Make your choice and press E.
E
↓
pm152e09.chp
mycom CPM 152
Calibration
Hauser
65