11.4
Service
Press the DIAG key and choose “Service”
→
“Specialist”.
Following you will be offered a choice of the
following groups:
• Simulation
• Internal data
• Factory reset
• Instrument check
• Special functions
Note:
The value displayed for the current
output and/or the contact status
displayed is immediately set in the
“Simulation” menu. If the value in the
window is changed, the current
output and/or the contact status is
changed at the same time.
The simulation facility is deactivated
by leaving the window.
11.4.1
Simulation
11.4.2
Internal data
Function
Selection
Setting the present
current value on current output 1
0.00 ... 22.00 mA
Setting the present
current value on current output 2
0.00 ... 22.00 mA
Setting the current contact statuses
(the number of contacts displayed depends
on the instrument equipment and
configuration)
Selection of the contacts with the
↑
and
↓
keys. Opening and closing of the
contact selected with the
→
key.
Simulation of measured pH value
1)
–2.00 ...16.00 pH / ±1500 mV / ±3000 %
Temperature simulation
1)
–20.0 ... 150.0 °C
1)
Current and contact outputs change according to the set configuration.
Function
Selection
Display of the device number
No selection
Display of the software version
No selection
Display of the hardware configuration in
several consecutive windows: slot allocation,
module, installation date, module serial
number
No selection
Order code
Alphanumeric input with the characters
0 ... 9 and a ... Z
Reset counter
No selection
Diagnostics