Additional module FCP1:
for second pH redox input, identical to
module FCP1 in the basic version
pH/mV pH/redox input
not connected
not connected
Ref.
Reference
PA
Potential matching
13
Pt 100, compensation cable
12
Pt 100, sensor cable
11
Pt 100, sensor cable
Additional module FCYK:
with 3 relays for controllers
and/or Chemoclean
89/90
Contact 2
91/92
Contact 3
93/94
Contact 4
Additional module FCXI:
with 2 binary input contacts and active
analog input (measuring transmitter supply)
21
Analogue current input plus
22
Analogue current input minus
81
Binary input 1 plus
82
Binary input 1 minus
83
Binary input 2 plus
84
Binary input 2 minus
Function of the binary inputs: see chapter 6.8.
Ref
13 12 11
Pt
100
fcp2con.cdr
pH/
mV
PA/
PM
Fig. 4.8
Connection of
module FCP1
Contact 2
Contact 3
Contact 4
89
90
91
92
93
94
fcykcon.cdr
Fig. 4.9
Connection of module
FCYK, non-Ex
21
22
81
82
83
84
mA
Binary inputs
fcxicon.cdr
Analogue input
Fig. 4.10
Connection of module
FCXI
Technical data
Analogue current input with measuring transmitter supply (for Autoclean system)
Measuring current range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 ... 22 mA
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20 V at 30 mA
Binary inputs . . . . . . . . . . . . . . . . . . . . Passive, external energy supply necessary
Power supply:
Terminal voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . max. 30 V, nom. 12 V
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . nominal
≈
2 mA
Isolation voltage . . . . . . . . . . . . . . . . . . . . . . . . Galvanically separated 276 V
eff
pm152e04.chp
mycom CPM 152
Installation
Hauser