11.2
Information list / logbook
11.2.1
Information list
Two information windows are displayed in
succession in the menu point “Information
list”. In window 1 you will find the last time the
instrument was started up, its name and
the number of control contacts. You can move
on to window 2, the “logbook”, by pressing
the E key.
11.2.2
Logbook
Window 2 contains a list of the last 30 times
the instrument has been operated, including
the date and time. The latest occurrence is
shown at the top of the list. When the list is
full, the last entry is deleted in each case.
The entries cover:
• all changes to the configuration
• all simulations (in general)
It is not possible to delete or change entries in
the logbook.
Fault in process
No.
Display
Remedy
E055
Display range measured value
under limit
Check measurement, control and
connection; if necessary, check
instrument and measuring cable with
simulator
E056
Display range measured value 2
under limit
E057
Display range measured value
over limit
E058
Display range measured value 2
over limit
E059
Temperature range under limit
E060
Temperature range 2 under limit
E061
Temperature range over limit
E062
Temperature range 2 over limit
E063
Current limited 0/4 mA output 1
Check the configuration in the “Current
outputs” menu; check measurement,
control and connections; if necessary
check instrument and measuring cable
with simulator
E064
Current limited 20 mA output 1
E065
Current limited 0/4 mA output 2
E066
Current limited 20 mA output 2
E067
Limit or setpoint 1 exceeded
E068
Limit or setpoint 1 exceeded
E072
Loss in compressed air pressure (only
with CPC 20)
Check compressed air supply
Warning
E080
Range for current output 1 too small
Increase the range in the “Current
outputs” menu
E081
Range for current output 2 too small
Diagnostics
mycom CPM 152
78