7.1
System Set-up
7.1.1
Parameter
Note:
When the parameter is changed, all
the data are reset to the settings
made at the factory (default settings).
→
Commissioning
→
System set-up
→
Parameter
☞
7.1.1
→
Pass codes
☞
7.1.2
→
Output relays
☞
7.1.3
→
General settings
☞
7.1.4
Function
Selection
Default
Operating mode
pH,
redox absolute (mV),
redox relative (%)
pH
Measurement type
(only with two-circuit instrument)
Single circuit measurement channel 1,
Single circuit measurement channel 2,
redundancy measurement
Redundancy
measurement
Electrode type
(only with operating mode pH)
Glass electrode pH 7.0
Glass electrode pH 4.6
Antimony
Glass
electrode
pH 7.0
PM connection, potential
matching
Without PM (potential matching)
(asymmetrical electrode connection)
With PM (potential matching)
(symmetrical, high-impedance
electrode connection)
with PM
Control parameters for 2nd
current output (only with
two-circuit instrument)
pH / redox channel 2
Diff.
∆
channel 1 – channel 2
Temperature (channel 1)
Temperature
PARAM
Instrument configuration
mycom CPM 152
32