9.3
Calibrate redox relative (%)
•
Data entry absolute
(
☞
9.3.1)
Entry of electrode offset in mV.
•
Data entry relative
(
☞
9.3.2)
Entry of two % calibration points
and to each of them the according
mV value.
•
Calibration absolute
(
☞
9.3.3)
The electrode offset is calculated out of the
difference between the actual measured
value and the known buffer value.
•
Calibration relative
(
☞
9.3.4)
One detoxified and one unchanged
sample serve as buffers.
•
50% end point calibration
(
☞
9.3.5)
The experimentally determined 50% end
point is used.
9.3.1
Data entry absolute (%)
9.3.2
Data entry relative (%)
Display
Description
Enter the mV value for the electrode offset.
Press the E key. The value entered takes
effect immediately.
Display
Description
Form two pairs of relative and absolute
redox values by entering appropriate figures:
Select line with “
↑↓
” keys.
Press “
→
” if you want to edit the selected
figure using the “
↑↓
” keys.
Confirm each edited value by pressing “E”.
Pressing the “E” key once more confirms the
whole new settings. The input will take effect
immediately.
↓→
↑
↓→
Calibration
mycom CPM 152
64
Hauser