Installation
Cerabar PMP71B HART
24
Hauser
Ordering information:
• Can be ordered via the Product Configurator
• Can be ordered as a separate accessory, part number 71102216
If the measuring device is ordered with a separate housing, the mounting bracket is
included in the delivery.
When mounting on a pipe, tighten the nuts on the bracket uniformly with a torque of at
least 5 Nm (3.69 lbf ft).
5.2.10 Assembling and mounting the separate housing
r
³
120 (4.72)
7
1
2
1
6
6
5
4
3
A0038728
Unit of measurement mm (in)
1
Housing mounted with housing adapter, included
2
Mounting bracket provided, suitable for pipe and wall mounting (for pipe diameters from 1 ¼" up to 2")
3
Locking screw
4
Plug
5
Pressure compensation
6
Cable with connection jack
7
In the separate housing version, the sensor is delivered with the process connection and cable already
mounted.
Assembly and mounting
1. Insert the connector (item 4) into the corresponding connection jack of the cable
(item 6).
2. Insert the cable with the socket (item 6) into the housing adapter (item 1) to the end
stop.
3. Tighten the locking screw (item 3).
4. Mount the housing on a wall or pipe with the mounting bracket (item 2). When
mounting on a pipe, tighten the nuts on the bracket uniformly with a torque of at
least 5 Nm (3.69 lbf ft). Mount the cable with a bending radius (r) ≥
120 mm (4.72 in).