11.6.1 List of diagnostic events
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
062
Sensor connection faulty
Check sensor connection
F
Alarm
081
Sensor initialization
faulty
1. Restart device
2. Contact service
F
Alarm
100
Sensor error
1. Restart the device
2. Contact Hauser
Service
F
Alarm
101
Sensor temperature
1. Check process temperature
2. Check ambient temperature
F
Alarm
102
Sensor incompatible error 1. Restart device
2. Contact service
F
Alarm
Diagnostic of electronic
242
Firmware incompatible
1. Check software
2. Flash or change main electronic
module
F
Alarm
252
Module incompatible
1. Check if correct electronic
modul is plugged
2. Replace electronic module
F
Alarm
263
Incompatibility detected
Check electronic module type
M
Warning
270
Main electronics
defective
Replace main electronics
F
Alarm
272
Main electronics faulty
1. Restart device
2. Contact service
F
Alarm
273
Main electronics
defective
1. Transfer data or reset device
2. Contact service
F
Alarm
282
Data storage inconsistent 1. Restart device
2. Contact service
F
Alarm
283
Memory content
inconsistent
1. Transfer data or reset device
2. Contact service
F
Alarm
287
Memory content
inconsistent
1. Restart device
2. Contact service
M
Warning
311
Electronic failure
Maintenance required!
1. Do not perform reset
2. Contact service
M
Warning
Diagnostic of configuration
410
Data transfer failed
1. Check connection
2. Retry data transfer
F
Alarm
412
Processing download
Download active, please wait
C
Warning
431
Trim required
Carry out trim
C
Warning
435
Linearization faulty
Check linearization table
F
Alarm
437
Configuration
incompatible
1. Restart device
2. Contact service
F
Alarm
438
Dataset different
1. Check data set file
2. Check device configuration
3. Up- and download new
configuration
M
Warning
441
Current output out of
range
1. Check process
2. Check current output settings
S
Warning