ASP station 2000 peristaltic
Commissioning
Hauser
99
!
Note!
Pressing the AUT key irreversibly deletes all the data saved up to this point in the datalogger.
6.4.3
Configuring the outputs
Menu structure:
Fig. 40:
Outputs menu structure
6.4.4
Programme selection
!
Note!
It is only possible to select the
switching and event programmes
in the 7-programme mode!
Power down
The output is switched when the controller is shut down (battery voltage < 10 V)
Sample
The output is switched on a sample.
Bottle
The output is switched on bottle replacement.
Ext. stop
The output is switched on an external stop.
Fill. end
The output is switched on exiting the sampling programme.
No flow
The output is switched if the device was unable to draw in any sampling medium (e.g.
blocked suction hose).
Error
The output is switched on error.
Suction
The output is switched on a sample during intake.
Overfill
The output is switched when a bottle is overfull.