ASP station 2000 peristaltic
Operation
Hauser
89
Fig. 24:
Multi-parameter probe calibration
The menu structure for calibration is identical for the individual sensors and appears as follows:
Fig. 25:
Menu structure for sensor calibration
If the calibration is successful, the message 'CALIBR. SUCCES.' If the measured values are outside
specific limits, the message 'VALUE OUTSIDE' is displayed and the user is then prompted to
'ACCEPT CALIBRATION' or 'ABORT'. If a sensor cannot be calibrated, it may have to be replaced.
If sensors are calibrated too often, this can have a negative effect on the accuracy. For this reason,
it is possible to reset the sensors to the factory setting. In the sensor calibration menu in question,
there is the item UNCAL. If this is confirmed with YES, the sensor is reset to its original value. The
message 'UNCAL OK' appears on the display.
Calibrating conductivity measurement
For calibrating the conductivity sensor, you can choose from the following output forms: specific
conductivity, conductivity in original form and salinity. One version is enough to calibrate the
conductivity sensor.
Act. value
Displays the current measured value of the sensor, the unit corresponds to the
setting made.
Uncal
Calibration values are reset to the factory setting when you confirm with YES.
Cal. value
For entering the value of the calibration solution used.
Calibrate
The calibration is confirmed by selecting ACCEPT.
Содержание ASP Station 2000 peristaltic
Страница 66: ...ASP Station 2000 Peristaltik Index 66 Endress Hauser W Warenannahme 8 Werkstoffe 63 ...
Страница 197: ...ASP Station 2000 péristaltique Index Endress Hauser 197 ...
Страница 198: ...Endress Hauser ...
Страница 199: ...Endress Hauser ...
Страница 200: ...www endress com worldwide BA176R 09 a3 04 06 51008239 FM SGML6 0 ProMoDo ...