ASP station 2000 peristaltic
Trouble-shooting
Hauser
119
Control signals not accepted or outputs
do not switch
Incorrect programme setting
Incorrect wiring
Electronics failure
Check programming
(see Section 6)
Check wiring
(see Section 4)
Replace controller (only by E+H
Service)
Battery cannot be charged
Wrong charger
Fuse defective
Only use E+H charger (see spare parts
list)
Replace fuse (only by E+H Service)
Sample not representative
Siphon
Connection not tight
Sampling hose drawing in air
Bottle not filling correctly
Distributor tap stuck
Wrong bottles moved forward
Distribution stops at bottle
Incorrect pump hose
Sensory mechanism faulty
See Section 3
Check tightness of hoses/connections
Uneven mounting
Tap incorrectly calibrated
Incorrect distribution selected in
operation
Incorrect distribution selected in
operation
Distributor defective, repair by E+H
Service
Incorrect distribution selected in
operation
Bottle clamp missing
Only use E+H pump hose
Replace sensory mechanism (only by
E+H Service)
No sampling
Connection not tight
Sampling hose drawing in air
Incorrect pump hose
Sensory mechanism faulty
Check tightness of hoses/connections
Only use E+H pump hose
Replace sensory mechanism (only by
E+H Service)
Problem
Cause
Remedy
Содержание ASP Station 2000 peristaltic
Страница 66: ...ASP Station 2000 Peristaltik Index 66 Endress Hauser W Warenannahme 8 Werkstoffe 63 ...
Страница 197: ...ASP Station 2000 péristaltique Index Endress Hauser 197 ...
Страница 198: ...Endress Hauser ...
Страница 199: ...Endress Hauser ...
Страница 200: ...www endress com worldwide BA176R 09 a3 04 06 51008239 FM SGML6 0 ProMoDo ...