Commissioning
ASP station 2000 peristaltic
104
Hauser
6.4.5
Creating a main programme
Menu structure:
Sampling types
Menu structure:
Fig. 48:
Sampling types menu structure
You can take samples proportional to time, quantity, flow or controlled by events (see also "Sampling
types").
Fig. 47:
Main programme menu structure
Programme name
The programme name can be changed as required with
the operating keys.
Time
Sampling is triggered after a settable time.
Quantity
Sampling is triggered after a specific measured flow rate.
!
Note!
For this function, a quantity signal must be connected to the analogue input or one of the
digital inputs of the sampler. To do this, the digital input must be configured using the
QUANTITY function.
Ext. sig.
Sampling is triggered by an external signal.
!
Note!
One of the digital inputs must be assigned and configured using the SAMPLE function.
Flow
Sampling is triggered after a settable time. The volume sampled is proportional to a
measured flow rate.
!
Note!
A quantity signal must be connected to the analogue input of the sampler for this function.