ASP station 2000 peristaltic
Maintenance and cleaning
Hauser
115
7
Maintenance and cleaning
7.1
Cleaning the device
Only clean the device when it is switched off. Use only a cleaning agent which you are sure will not
damage the mechanical and electrical equipment.
For the cabinet body, we recommend a stainless steel cleaner. For any parts conveying media, use
water or soap. Thorough, regular cleaning of the parts conveying media is essential for reliable
operation.
!
Note!
You can mount and disassemble all the parts which convey media easily and without tools.
7.2
Cleaning parts which convey media
7.2.1
Cleaning the hose pump
#
Warning!
• If you open the hose pump while the sampler is in operation, your fingers are at risk of being
crushed!
• For this reason, make sure that the sampler has been taken out of service (activate "Off" switch)
before you open the hose pump. Once the LC display goes out, the sampler is out of service.
• Secure the sampler against unintentional start-up whilst you carry out work on the open hose
pump.
"
Caution!
Please remove the pump hose if taking the device out of service for an extended period (> 2
months)!
Replacing the pump hose
During operation, the pump hose is subject to natural wear and tear. For this reason, check the
pump hose for leaks at regular intervals. If the pump hose is leaking, replace it as per the separate
installation instructions. The installation instructions are provided in the same packaging as the
pump hose.
Fig. 62:
Cleaning parts which convey media
1.
Hose pump: clean as described in
2.
Inner suction hose: unscrew from
hose pump and hose fitting. Rinse
with soap and water.
3.
Tap: pull forwards and out and
remove cover. Rinse with soap
and water.
4.
Distribution pans: pull forwards
out of the frame. Rinse with soap
and water.
5.
Remove the bottle trays. Clean
the bottles and bottle trays with
soap and water.
Содержание ASP Station 2000 peristaltic
Страница 66: ...ASP Station 2000 Peristaltik Index 66 Endress Hauser W Warenannahme 8 Werkstoffe 63 ...
Страница 197: ...ASP Station 2000 péristaltique Index Endress Hauser 197 ...
Страница 198: ...Endress Hauser ...
Страница 199: ...Endress Hauser ...
Страница 200: ...www endress com worldwide BA176R 09 a3 04 06 51008239 FM SGML6 0 ProMoDo ...