ASP station 2000 peristaltic
Commissioning
Hauser
97
6.4
Device configuration
!
Note!
The figures show the menu structures with the display and picklist for the different configuration
parameters. Below the figures are tables containing the corresponding function descriptions for the
parameters.
6.4.1
Configuring the inputs
You can assign the inputs when configuring in the "Quick Setup" (
Digital inputs
Menu structure:
Fig. 37:
Digital input menu structure
Sampling
The input signal triggers a sample.
Bottle
The input signal triggers a switch to the next empty bottle.
Quantity
The input signal is a pulse signal from a volume meter; alternative to an analogue quantity
signal (0/4...20 mA).
Ext. stop
The input signal interrupts any programmes running; the programmes will be continued
when the signal disappears.
Event
The input signal triggers an "event sample". For example, the input signal may be switched
when a limit value is infringed; a separate bottle can be filled in an event sample.
Report
A message appears in the display showing the date and time (e.g. function error of
flowmeter). You must acknowledge the message; the sample programme is not
interrupted.
Switch
The input signal triggers a change to the switching programme.
Содержание ASP Station 2000 peristaltic
Страница 66: ...ASP Station 2000 Peristaltik Index 66 Endress Hauser W Warenannahme 8 Werkstoffe 63 ...
Страница 197: ...ASP Station 2000 péristaltique Index Endress Hauser 197 ...
Страница 198: ...Endress Hauser ...
Страница 199: ...Endress Hauser ...
Страница 200: ...www endress com worldwide BA176R 09 a3 04 06 51008239 FM SGML6 0 ProMoDo ...