ASP station 2000 peristaltic
Commissioning
Hauser
105
Distribution
Menu structure:
Fig. 49:
Distribution menu structure
You can programme a bottle change by time, number or an external signal:
Start-stop operation
Menu structure:
Fig. 50:
Start-stop operation menu structure
You can start the sampling programme either immediately by pressing the AUT key, or at a settable
time. You can specify the sampling programme stop as follows:
• Programme end: the unit stops sampling automatically when it has run through the set
programme.
"
Caution!
Risk of overfilling!
• No: the unit runs through the programme in a loop. Do not forget to empty the bottles.
• Time: the sampling programme stops at a settable time.
You can choose between continuous operation and operation at different intervals for the operating
mode.
• Day: operating time at two programmable times per day.
• Week: operating time on three programmable days per week.
• Interval: operating at specific time intervals.
Time
After a settable time, the tap switches to the next empty bottle.
Number
After a set number of samples has been taken, the tap switches to the next empty bottle.
Ext. sig.
The tap switches to the next empty bottle when an external signal is present.
!
Note!
One digital input must be assigned and configured using the BOTTLE function.
Содержание ASP Station 2000 peristaltic
Страница 66: ...ASP Station 2000 Peristaltik Index 66 Endress Hauser W Warenannahme 8 Werkstoffe 63 ...
Страница 197: ...ASP Station 2000 péristaltique Index Endress Hauser 197 ...
Страница 198: ...Endress Hauser ...
Страница 199: ...Endress Hauser ...
Страница 200: ...www endress com worldwide BA176R 09 a3 04 06 51008239 FM SGML6 0 ProMoDo ...