Operation
ASP station 2000 peristaltic
82
Hauser
5
Operation
5.1
Quick operation guide
Fig. 11:
Quick operation guide
5.1.1
"Man" key
Fig. 12:
"Man" key
The "Man" key is used to trigger immediate sampling regardless of whether a programme has been
started or not. The sampling process starts immediately once the key has been pressed.
"
Caution!
The sample is discharged at the point where the tap is currently located. If the tap has been adjusted
or if a programme has not been started after switch-on, the tap can be located between two
discharge positions.
5.1.2
"Off" key
The "Off" key has different functions depending on whether a programme has been started or not.
Programme has not yet been started
If a programme has not been started, pressing the "Off" key simply switches off the device. This
means that the "Man", "Aut" or "->" keys are no longer available.