ASP station 2000 peristaltic
Operation
Hauser
85
The messages that appear on the display can be acknowledged by pressing the "OK" key.
5.1.4
"Stat" key
Fig. 16:
"Stat" key
The bottle statistics of the sampler are displayed once the "Stat" key is actuated. Statistics are
displayed for each individual bottle since the programme start. In this way, the user can draw
conclusions on the last samplings. The statistics are deleted for the following events:
• Programme start
• 1st bottle is reached, if "Programme end: no" is set as the programme end in the programme
settings.
The statistics are displayed as follows:
Fig. 17:
Bottle statistics
• The bottle number is displayed in the first column (no.).
• The second column (smp.) indicates how often sampling was triggered per bottle.
• The third column (n.smp) indicates the number of times a sample was not taken even though
sampling was triggered. This can happen if the maximum permitted filling volume of the bottle
was achieved but samples should still be deposited into this bottle, for example. In this instance,
the text "Overfill security" appears on the display when the programme is running.
• The fourth column (n.bot) indicates how often sampling was aborted since no medium, or not
enough medium, could be drawn into the dosing beaker to wet the LF1 probe.
Содержание ASP Station 2000 peristaltic
Страница 66: ...ASP Station 2000 Peristaltik Index 66 Endress Hauser W Warenannahme 8 Werkstoffe 63 ...
Страница 197: ...ASP Station 2000 péristaltique Index Endress Hauser 197 ...
Страница 198: ...Endress Hauser ...
Страница 199: ...Endress Hauser ...
Страница 200: ...www endress com worldwide BA176R 09 a3 04 06 51008239 FM SGML6 0 ProMoDo ...