PlasmaQuant PQ 9000 (Elite)
Specification
81
Supply voltage
Frequency
230 V ±10%
50/60 Hz
Mains fuse installation in
the building
32 A
Typical average power con-
sumption
Base device: 4500 VA
Overvoltage category
II according to DIN EN 61010-1
Degree of contamination
2 according to DIN EN 61010-1
Protection class
I
Protection type
IP 20
Fuse number
Type
Protected circuit
S1
10 A NFC 10x38 gG AC
400 V
Spectrometer
S2
6 A NFC 10x38 gG AC 400
V
Tubular heating generator
S3
25 A NFC 10x38 gG AC
400 V
Power supply unit generator
Monitoring of
Plasma compartment door closure
Torch working position
Cooling
Waste air
Argon supply
Plasma (optical monitoring)
Recommended gas
Inlet pressure
Consumption (total)
Argon 4.6 or superior
Permitted components:
Oxygen
≤ 3 ppm
Nitrogen
≤ 10 ppm
Hydrocarbons
≤ 0.5 ppm Hu-
midity
≤ 5 ppm
6 bar
13-21 L/min
Additive gas (optional)
Oxygen 4.5 or superior
6 bar
Max. 0.04 L/min
According to DIN ISO 9022-2:2003/01
Corrosion protec-
tion
The device is corrosion-proof against the samples used in the analysis
Operating tempera-
ture
+15°C to +35°C, at best +20°C to +25°C
as constant as possible during measuring; air conditioning is recom-
mended
Humidity
20 to 90% at 20°C
Power supply
Instrument fuses
Safety circuits
Gas supply
Environmental conditions