Specification
PlasmaQuant PQ 9000 (Elite)
80
10
Specification
10.1
Technical data
10.1.1
PlasmaQuant PQ 9000 data
Monochromator
Echelle grating double-monochromator with a focal length of F=400
mm and variable gap; pre-monochromator with quartz prism, wave-
length selection via additional reflected neon radiator
Wavelength range
160 – 900 nm
Wavelength accuracy < 0.4 pm
Spectral resolution
0.002 nm at 200 nm; 0.006 at 400 nm; 0.009 at 600 nm
FWHM values
≤ 3.5 pm for As 193.696 nm, P 213.618 nm and Cd 228.022 nm
Resolution
1:145,000
Grating
Mechanically ruled grating, 79 grooves/mm, blaze angle of 76°
Optical bank
encapsulation pho-
tometer
Optics in modular construction on compact cast base plate for stability
and endurance;
protection against moisture, exhaust fumes and chemical environ-
mental impacts
Detector
Two-dimensional FFT backside illuminated CCD with high quantum
efficiency and increased UV sensitivity
Monochromator
Echelle grating double-monochromator with a focal length of F=400
mm and variable gap; pre-monochromator with quartz prism, wave-
length selection via additional reflected neon radiator
Wavelength range
160 – 900 nm
Wavelength accu-
racy
< 0.4 pm
Spectral resolution
0.006 nm at 200 nm
Resolution
1:70,000
Grating
Mechanically ruled grating, 79 grooves/mm, blaze angle of 63°
Optical bank
encapsulation pho-
tometer
Optics in modular construction on compact cast base plate for stability
and endurance;
protection against moisture, exhaust fumes and chemical environ-
mental impacts
Detector
Two-dimensional FFT backside illuminated CCD with high quantum
efficiency and increased UV sensitivity
Emission
Counts (ct)
Intensity
Counts/seconds (ct/s)
Concentration
Value range: 5 characters (0.0001 to 99999), unit freely selectable
Spectrally resolved
Specters of 20 pixels to a maximum of 200 pixels in width
Optical system Plasma-
Quant PQ 9000 Elite
Optical system Plasma-
Quant PQ 9000
Display modes
Signal evaluation