Maintenance and care
PlasmaQuant PQ 9000 (Elite)
68
Figure 30 Replacing the fuses of the autosampler
6.4
Maintenance of the recirculating chiller - Changing the cooling water
CAUTION
Biocide! Observe the safety data sheet of the biocide used.
NOTICE
Observe the cooling water preparation!
The cooling water additive for recirculating chillers (order no. 418-13-410-540) must
be added to the cooling water. This is the only way to effectively avoid damages to the
device due to corrosion or biological contamination.
Warranty claims are excluded if damages to the PlasmaQuant PQ 9000 are due to the
fact that the cooling water additive was not used in accordance with these instructions!
The cooling water must be replaced every year.
10 L of distilled/deionized water
Biocide/corrosion protection set for chiller (order no. 418-13-410-540)
A suitable appropriately sized vessel made of glass, plastic or stainless steel for mix-
ing the coolant
A bucket to catch the drained coolant
1.
Dissolve both bottles of the biocide / corrosion protection set in 10 L of water.
2.
In ASpect PQ, start the wizard for changing the coolant and follow the instructions in
the wizard:
Switch off the recirculating chiller.
At the recirculating chiller remove the inlet connection and hold the tube in the re-
ceptacle (bucket).
Switch on the recirculating chiller again and allow to run until the coolant flow ends
and only spray comes out.
Reconnect the tube to the inlet connection on the recirculating chiller.
Unscrew the sealing cap from the filler opening of the tank and insert the funnel.
Pour the prepared coolant into the tank until the maximum level mark is reached.
Required equipment: