multi N/C pharma UV
Fault removal
89
24
Optics error, analog values outside of range
Cause
Remedy
the analog values of the detector are outside the
working range
check the quality of the carrier gas
NDIR detector faulty
inform Service
40
no connection to the syringe pump
Cause
Remedy
no communication between analyzer and syringe
pump
initialize analyzer
switch off PC, switch back on and initialize
analyzer
81
UV cover open
Cause
Remedy
contact at UV cover not closed (e. g. after replacing
the UV module)
close cover
84
Communication error in temperature controller
Cause
Remedy
communication error
inform Service
111
Rotator error
Cause
Remedy
Drive incorrectly positioned, e.g. jammed
Drive faulty
Initialize analyzer
If the error cannot be corrected, contact
Service
112
Swivel drive error
Cause
Remedy
Drive incorrectly positioned, e.g. jammed
Drive faulty
Initialize analyzer
If the error cannot be corrected, contact
Service
113
Lifting drive error
Cause
Remedy
Drive incorrectly positioned, e.g. jammed
Drive faulty
Initialize analyzer
If the error cannot be corrected, contact
Service
114
Rack detection error
Cause
Remedy
Sample tray not positioned correctly
Position the sample tray again and make
sure it clicks into place
Initialize analyzer
115
Wrong rack
Cause
Remedy
Wrong sample tray set in the software
Check settings in the software (see section
5); if necessary, set a different sample tray