Operation
multi N/C pharma UV
64
6.4.5
Managing calibration data
Print the calibration report as follows:
1.
In the window C
ALIBRATION
–
C
ALIBRATION SETTINGS
enable the option
USE CALIBRATION
.
2.
Define the scope of printing under the menu P
RINT OPTIONS
:
Print calibration graph and/or
print individual integers for each calibrated channel
3.
Start the printout with the menu command C
ALIBRATION
R
EPORT
P
RINT
.
Calibration data are exported via the menu D
ATA EXPORT
in the window C
ALIBRATION
–
C
ALIBRATION SETTINGS
. You have the following options for exporting calibration data:
CalibrationReport into an export file
The calibrationReport (with the extension *.ajc) is saved in the export directory
..\C
ALIBRATION
.
Export into a CSV file (*.csv)
The CSV file is saved in the preconfigured directory (default ...\multiWin\CSV).
The directory is selected in the window O
PTIONS
tab F
ILES AND DIRECTORIES
(main
window menu command C
ONFIGURATION
E
DIT OPTIONS
).
Export to clipboard
1.
In the main window open the menu command D
ATA
E
VALUATION
C
ALIBRATION
R
EPORT
.
2.
In the window S
ELECTION
C
ALIBRATION
R
EPORT
select the calibration report.
In the window S
ELECTION
C
ALIBRATION
R
EPORT
filters can be set, if necessary, and the
records sorted by clicking on the respective header.
Fig. 35 Window
CALIBRATION REPORT SELECTION
Printing calibration data
Exporting calibration files
Reopening a calibration
report
Содержание Analytik Jena multi N/C