multi N/C pharma UV
Fault removal
87
8.2
Error messages in multiWin
Error code
Error message
VERS
Communication error - incorrect command set between PC and device!
Cause
Remedy
the internal and external program versions do
not match
update the internal and external program
VERS1
Communication error - analyzer
Cause
Remedy
analyzer not switched on
switch on analyzer
multiWin started too early
only start multiWin after 30 sec
analyzer not connected to PC
check connection between analyzer and PC
incorrect COM port set at the external computer
check set interface at the external computer, if
necessary select different interface in multiWin
with menu command C
ONFIGURATION
INTERFACE
-6
Analysis device is busy
Cause
Remedy
analyzer > 10 min in the busy state
initialize analyzer
-5
communication error - analyzer STAT, MESS, STEP or INIT
Cause
Remedy
communication error
initialize analyzer
-4
Communication error - analyzer
Cause
Remedy
communication error
check interface cable
initialize analyzer
-3
-2
-1
command from the analyzer CRC error
CRC error
invalid command from the analyzer
Cause
Remedy
communication error
initialize analyzer
1
2
3
4
5
6
Incomplete command from the PC
PC command without STX
PC command without *
PC command CRC error
PC command invalid command
PC command invalid MESS command
Cause
Remedy
faulty connection between internal and external
program
initialize analyzer