Connecting add-on devices
multi N/C pharma UV
50
4.
Connect the table power supply cable on the low voltage side to the rear of the
autosampler. Do not connect the power supply to the grid yet.
5.
Connect the data cable supplied to the "Sampler" port on the rear of the analyzer.
6.
Connect the other end of the data cable to the RS 232 port of the autosampler.
7.
Connect the waste hose supplied to the wash cup of the autosampler and to a
suitable waste container or drain.
8.
Fit the wash cup to the autosampler.
9.
Place the sample tray onto the space provided.
Note the positioning of the tray. The label has to be legible if you face the front of
the device. The two centering pins (black plastic) on the contact surface of the
autosampler have to protrude into the drill holes in the tray floor.
10.
Insert piercing cannulas and holding-down clamps into the autosampler arm.
For NPOC measurements with parallel
purging:
Insert 1 cannula into each of the two
positions in the cannula holder
For NPOC measurements with non-
parallel purging:
Insert both cannulas into the left position
in the cannula holder
11.
Clamp the two cannulas high enough in the holder to prevent them dipping into
the tubes (basic position).
12.
Connect the two connection hoses to the analyzer to the cannulas:
Hose no. 6 = sample aspiration hose
Hose no. 10 = purging hose for NPOC measurements
Release the upper Fingertight connection of the cannula.
Guide the hose through the banjo bolt.
Slide the conical nipple with the conical side towards the banjo bolt onto the
hose. The conical nipple and hose must be flush.