Maintenance and care
69
ATTENTION
To carry out regular inspections and maintenance tasks always ensure that the doors
and the left side wall of the analyzer are freely accessible.
ATTENTION
Ensure that all connections are gas-tight again after servicing:
Do not insert the Fingertight screw connections twisted!
Tighten all screw connections finger-tight!
Check the system for leaks (see section "Checking the system for tightness" p. 84).
7.2
Adjustment and setup tasks
7.2.1
General notes for adjusting the autosampler
During adjustment, the cannulas to the sample tray are adjusted so that they are
optimally immersed into the sample tubes and/or wash cups.
An adjustment of the sampler is necessary:
before the first start
after each change in the size of the sample tray
during recommissioning after transport or storage
Adjusting the AS 10 and AS 21 autosamplers is described in the "Connecting add-on
devices" section, p. 38.
7.2.2
Adjusting the AS vario autosampler
ATTENTION
The cannulas can bend! Before adjusting the sampler undo the screw connections of
the sample aspiration and purging cannulas!
For NPOC measurements, the immersion depth for automatic acidification (z position)
depends on the immersion depth in position 1. Adjust the cannula in position 1 and
check the adjustment by a test measurement. Make sure the cannula perforates the
cover (for example aluminum foil) but does not enter the sample liquid when
delivering the acid.
1.
Start the multiWin software and wait for device initialization.
2.
Select the
I
NSTRUMENT
S
AMPLER ALIGNMENT
menu command to open the window
with the same name.
3.
In the P
LEASE SELECT POSITION NEEDING ADJUSTMENT
group, select
NEEDLE
from the list
field.
The autosampler arm will move over the adjustment points on the sample tray.