Disposal
10.1
Waste water
Waste water arises during the ongoing analysis operation of the multi N/C pharma UV.
This water contains sulfuric acid, diluted phosphoric acid and sample, depending on
the measurement mode.
The neutralized (if necessary) waste must be brought to the appropriate waste
disposal center for correct disposal according to the appropriate legal guidelines.
10.2
Halogen trap
The halogen trap contains copper. Contact the responsible institution (authority or
waste disposal company). There you will receive the information regarding recycling or
disposal.
10.3
Analyzer
At the end of its service life the multi N/C pharma UV and all its electronic components
must be disposed of as electronic waste in accordance with the applicable regulations.
10.4
UV module
The UV module is equipped with a low-pressure mercury vapor lamp. For disposal,
remove the UV module from the analyzer (see section "Removing the UV module").
Dispose of the UV module in accordance with the national regulations for lamps that
contain mercury.
10.4.1
Removing the UV module
WARNING
Lethal electrical voltages may occur in the UV module of the analyzer! Touching the UV
module with the device switched on can be fatal!
Before opening the right side panel, press the main switch to turn off the analyzer and
remove the mains plug from the power outlet!
CAUTION
There is a risk of burning! The UV reactor remains hot immediately after switching off
the analyzer!
Wait at least 30 minutes for the UV reactor to cool down before removing the UV
module!