Connecting add-on devices
multi N/C pharma UV
46
1 Cannulas
2 Autosampler arm
3 Sample tubes
4 Sample tray
5 Grid connection and connection
to the analyzer (hidden)
6 On/Off switch
7 Fastening screws
Fig. 22 Layout of the AS 10 autosampler
1.
Plug cable on the low voltage side of the table power supply included in the
delivery into the connection at the bottom of the autosampler and connect the
power supply to the grid.
Make sure the AS 10 is switched off. (The green LED of the On/Off must be off.)
Connect the autosampler to the analyzer with the interface cable (port on the
bottom of the autosampler and "sampler" port on the rear of the analyzer).
2.
Screw the sampler with the two fastening screws to the right-hand side of the
equipment.
3.
Place a sample tube into position 1 of the sample tray.
4.
Insert the cannulas into the autosampler arm. Manually adjust the height of the
cannulas so that the cannula tips protrude 1 – 2 cm over the tube edge and cannot
touch the tubes when the autosampler arm is moving.
5.
Secure the cannulas by slightly tightening the screw.
6.
Switch on the AS 10 at the On/Off switch.
7.
Check the configuration in the multiWin program via the
I
NSTRUMENT
S
YSTEM
INFORMATION
menu command in the
SET
-
UP INFO
window. If necessary, modify the
configuration:
Exit multiWin.
In the Windows user interface, start the
SET
-
UP TOOL
under S
TART
P
ROGRAM
F
ILES
MULTI
W
IN
MULTI
W
IN SET
-
UP TOOL
.
Select the sampler type in the S
AMPLER
list.
Exit the
SET
-
UP TOOL
with [C
REATE
].
Start the multiWin program and select the
C
ONFIGURATION
E
DIT OPTIONS
menu
command to open the O
PTIONS
window on the
A
NALYZER COMPONENTS
tab. In the
Layout
Installation at the
analyzer