Connecting add-on devices
multi N/C pharma UV
44
In the P
LEASE SELECT POSITION NEEDING ADJUSTMENT GROUP
, select P
OSITION
1.
Click [P
OSITION
1
ADJUST
].
The autosampler arm now lowers the cannulas into the tube in position 1.
If necessary, increase or decrease the z values. Click [P
OSITION
1
ADJUST
] again
after each change to verify the change.
Once adjustment is complete, click [S
AVE
] to close the window.
By mounting a special cannula holder to the autosampler arm the autosampler can be
retrofitted for the "parallel purging" function.
Fig. 21 Cannula holder for parallel purging
1 Aspirating cannula
2 Purging cannula
3 Fixing screw on the spacer
4 Spacer
5 Clamping of the aspirating cannula
6 Clamping of the purging cannula
7 Cannula holder for "parallel purging"
8 Screw connection of the cannula holder
9 Clip for the attachment of the hoses
1.
Screw the cannula holder (7) to the autosampler arm if it is not preinstalled.
2.
Push the spacer (4) onto the two cannulas. Lightly secure the spacer below the
hose connections with the fixing screw, so that it cannot move.
3.
Insert the cannulas into the cannula holder according to the figure and attach them
only lightly using the knurled head screws.
4.
Place two sample cups into positions 1 and 2 of the sample tray under the two
cannulas.
5.
Manually adjust the height of the cannulas so that the cannula tips protrude 1 –
2 cm over the tube edge and cannot touch the tubes when the autosampler arm is
moving.
6.
Adjust the immersion depth of the cannulas in the tube (z direction) in multiWin.
Select the I
NSTRUMENT
S
AMPLER
A
LIGNMENT
menu command to open the
window with the same name.
In the P
LEASE SELECT POSITION NEEDING ADJUSTMENT GROUP
, select P
OSITION
1.
Retrofit to the function
"parallel purging"