Fault removal
multi N/C pharma UV
88
7
8
9
COM 2 not found
COM 3 not found
COM 4 not found
Cause
Remedy
problems with internal hardware
switch analyzer off/on
10
Gas pressure error
Cause
Remedy
counterpressure in the analyzer system too great:
carrier gas supply is automatically interrupted to
protect the analyzer; flow indication MFC (IN) approx.
0 ml/min
search for and replace component causing
the gas pressure error
water trap clogged
renew the water traps and check whether
the gas pressure error occurs again
see also p. 93
12
Incorrect version number
Cause
Remedy
the versions of multiWin and the software of the
internal computer do not match
update as appropriate
13
No connection to the autosampler
Cause
Remedy
sample not switched on
connection cable not connected or faulty
switch on sampler and initialize analyzer
check connection cable
15
Flow error in the window System state under gas flow
Flow leak shown in red and
in the window Measurement: values of IN and OUT shown in red
Flow indication IN: 140 ml/min
Flow indication OUT: < 130 ml/min
Cause
Remedy
system is leaking
UV reactor faulty (fractures at the connections)
TIC condensate faulty (fractures at the connections)
TIC condensate container connections leaking
connections at the water trap system loose (after
installing water traps, installing halogen trap)
aerosol/water trap clogged
hose pump leaking
check glass components, replace if faulty
FAST connector at TIC condensate
container, water traps
replace water traps (see error
aerosol/water trap clogged)
check hose pump, replace pump hose if
necessary
initialize analyzer
20
21
22
26
no connection to optics (NDIR)
CRC error optics
status error optics
optics error; incorrect command return
Cause
Remedy
communication error
initialize analyzer
NDIR detector faulty
inform Service