106
Profibus Communications
Diagnostic Bits
Signal
Board 1 / 2
GSD Descriptor
Help
40 / 72
SN 1/2 Board
Unknown (Fault)
The analyzer automatically recognizes the sensor board on power up. Sensor
board unknown implies the sensor board is new and not supported or there is a
bad connection between the sensor board and the main board.
1. Check ribbon cable connections at the main board and sensor board.
2. Inspect ribbon cable for damage and replace if necessary.
3. Replace sensor board.
41 / 73
SN 1/2 HW SW
Mismatch error
(Fault)
The main board software does not match the sensor board software.
1. Check ribbon cable connections at the main board and sensor board.
2. Inspect ribbon cable for damage and replace if necessary.
3. Replace sensor board.
42 / 74
SN 1/2
Incompatible
Error (Fault)
The sensor board software is not supported by the main board. Either the sensor
board software is too old or the main board software is too old. Replace the main
board with one compatible with the sensor board. Call the factory for assistance.
The main board must be replaced at the factory.
43 / 75
SN 1/2 Not
Detected
The sensor board is no longer being detected.
1. Check ribbon cable connections at the main board and sensor board.
2. Inspect the ribbon cable for damage and replace if necessary.
3. Replace the sensor board.
44 / 76
SN 1/2 Not
Communicating
(Fault)
The analyzer automatically recognizes the sensor board on power up. Sensor not
communicating implies there is a bad connection between the sensor board and
the main board or there is a bad component on the sensor board.
1. Check ribbon cable connections at the main board and sensor board.
2. Inspect the ribbon cable for damage and replace if necessary.
3. Replace the sensor board.
45 / 77
SN 1/2 CPU error
(Fault)
The sensor board software is corrupted.
1. Try to reset the analyzer by turning the power off then on again.
2. If cycling the power does not help, replace the sensor board.
46 / 78
SN 1/2 Factory
Data Error (Fault)
The factory-entered EEPROM data on the sensor board is corrupted.
1. Try to reset the analyzer by turning the power off then on again.
2. Replace the sensor board.
47 / 79
SN 1/2 ADC Error
(Fault)
There is a bad hardware component on the sensor board. Replace the sensor
board.
50 / 82
SN 1/2 Hardware
Error (Fault)
Hardware error means there is a missing or bad hardware component on the sen-
sor board. Replace the sensor board.
54 / 86
SN 1/2 User Data
Error (Fault)
The user-entered EEPROM data on the sensor board is corrupted.
1. Try to reset the analyzer by turning the power off then on again.
2. Replace the sensor board.
55 / 87
SN 1/2 EEPROM
Write Error (Fault)
There is a bad CPU on the sensor board.
1. Try to reset the analyzer by turning the power off then on again.
2. Replace the sensor board.
104 / 136
SN 1/2 Need
Factory Cal
(Warning)
The sensor board was not calibrated at the factory. Call the factory.
Section 12: Profibus Communications
Instruction Manual
April 2017
LIQ-MAN-56
Table 12-14. Signal Board Diagnostic Bits
Содержание Rosemount 56
Страница 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Страница 2: ......
Страница 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Страница 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Страница 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Страница 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Страница 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Страница 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Страница 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...