Section 7: PID Control
Instruction Manual
April 2017
LIQ-MAN-56
60
PID Control
Basic Definitions
•
Output
– Select the analog output (1 through 4) to be configured for PID control.
•
Analog/PID/Simulate
– Choose PID
•
Assign
– Select the Measurement to be controlled. Note: This measurement can also
be a 4-20 mA signal input brought in by the flow/ 4-20 mA board.
•
Range
– Select either 0-20 mA or 4-20 mA range, depending on the signal range used
by the final control element, e.g. a pump or valve.
•
Select
Next
to go to the PID Setup parameters.
7.2.3
PID Setup Parameters
The PID control setup window contains the PID control tuning parameters.
Also note that the upper portion of the screen shows the measurement chosen for control
(PV), and the control output in mA and % Output. This makes it possible to observe the
primary variable (PV) and the control output, in terms of percent and milliamps, while tuning
PID control.
PID Control Parameters: Basic Definitions
•
Setpoint
– Select the desired setpoint.
•
URV
– The value of PV (in the above example, 14.00 pH) at which the control will be 20
mA or 100% output.
•
LRV
-- The value of PV (in the above example 0.00 pH) at which the control will be 0 or 4
mA or 0% output.
Содержание Rosemount 56
Страница 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Страница 2: ......
Страница 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Страница 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Страница 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Страница 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Страница 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Страница 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Страница 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...