Instruction Manual
Appendix 1: HART and Device Variables
LIQ-MAN-56
April 2017
Appendix 1
117
pH Device Variables
Device Variable Name
Assignable to Dynamic
Variables
Variable Range
Sensor 1 / 2 Measurement Type
pH (1)
PV, SV, TV or QV
0 to 14 pH
ORP (2), Redox (3)
PV, SV, TV or QV
-1500 to 1500 mV
Ammonia (4), Fluoride (5), Custom ISE (6)
PV, SV, TV or QV
0 to 1000 ppm
0 to 1000 ppb
0 to 1000 mg/L
0 to 1000 µg/L
0 to 100 %
Measurement Parameters
Sensor 1 / 2 Temperature
PV, SV, TV or QV
-15 to 200 °C
5 to 360 °F
Sensor 1 / 2 Temperature resistance
TV or QV
0 to 100000 ohm
Sensor 1 / 2 mV input
PV, SV, TV or QV
-750 to 750 mV
Sensor 1 / 2 Glass impedance
PV, SV, TV or QV
0 to 2000 MΩ
Sensor 1 / 2 Sensor impedance
PV, SV, TV or QV
0 to 2000 MΩ
Sensor 1 / 2 Reference impedance
PV, SV, TV or QV
0 to 10000 kΩ
Appendix 1: HART and Device Variables
Содержание Rosemount 56
Страница 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Страница 2: ......
Страница 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Страница 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Страница 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Страница 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Страница 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Страница 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Страница 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...