Profibus Communications
Section 12: Profibus Communications
Instruction Manual
April 2017
LIQ-MAN-56
12.3.2
Profibus Faults and Warnings
Faults
The 56 displays the following hardware faults associated with problems related to the
Profibus board and its connection to the main board:
•
Profibus board mismatch -
The revision of the installed Profibus board is not compati-
ble with the main board of the transmitter. Note the revisions of the Profibus board and
the main board of the transmitter and contact the factory.
•
Self-test failed -
At power up, the Profibus board conducts several internal self tests. At
least one of them failed.
1. Try to reset the analyzer by cycling the power.
2. Replace the Profibus board
•
Internal communication failed -
The main board detects the Profibus board, but the
Profibus board is not communicating with the main board.
1. Try to reset the analyzer by cycling the power.
2. Check the interconnecting cable between the Profibus and main board.
3. Replace the Profibus board.
Warnings
Warnings related to cable problems and software compatibility are also displayed, if Profibus
cable diagnostics are enabled:
•
Cable shorted or open -
No Profibus cable is connected, or the A and B leads of the
cable are shorted.
•
Check Profibus Cable -
The A and B leads in the Profibus cable august be reversed. If
standard Profibus cable is used, the green lead should be wired to position A and the red
lead to position B. The bus august not have a terminator, or it august not be turned on.
Check to see that a terminator is present and turned on.
•
Profibus function limited -
The Profibus board has new features that the main board
software does not support. If this is the case, the minor revision number of the Profibus
board will be greater than the minor revision number of the main board. Replace the
main board.
Details on these faults and warnings are accessed like any other fault or warning in the 56:
Press the INFO key and then select “Faults – fix now” or “Warnings – fix soon.”
LED Diagnostics
The Profibus board itself displays diagnostic information via an LED, which is located on the
upper right side of an installed board:
•
Slow blink –
No cable is present or the connection is shorted
•
Slow, even blinking –
reversed A and B cable connection
•
Quick flashing –
Profibus message detected
•
1 second blinking –
self-test failure
•
Off –
no power or a defective or missing ribbon cable
Содержание Rosemount 56
Страница 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Страница 2: ......
Страница 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Страница 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Страница 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Страница 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Страница 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Страница 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Страница 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...