7. Enabling pH sensor diagnostics
To configure the pH measurement board, access the Program screen by pressing
ENTER/MENU from the main screen and
then select the Measurement tab.
1. To Select a Measurement type, se-
lect from: pH, ORP, Redox, Ammo-
nia, Fluoride, and Custom ISE and
press ENTER/MENU.
2. To program the Preamp location, se-
lect Analyzer or Sensor/JBox, and
press ENTER/MENU.
3. To Override the default input
filter, enter 0-999 seconds and press ENTER/MENU.
4. To program Reference Impedance. Select Low or High and press ENTER/MENU.
5. To choose the wiring scheme, Select Normal or Reference to Ground and press
ENTER/MENU.
6. To program the display resolution, Select 0.01 pH or 0.1 pH and press ENTER/MENU.
7. To enable pH sensor diagnostics, select the “pH diagnostic setup” tab under Program-
ming. Select Sensor 1 or Sensor 2. Select NEXT. Select On under sensor diagnostics to
enable pH diagnostics.
NOTE:
pH sensor diagnostics must be enabled to include diagnostic values such as Glass
Impedance and Reference Impedance in EXCEL data log sheets after data download to USB.
Enabling pH sensor diagnostics also allows assignment of Glass Impedance and Reference
Impedance to the two-dimensional on-screen process graph accessible under “Display
Setup/View Graph”.
6.3
ORP Measurement Programming
The section describes how to configure the 56 analyzer for ORP measurements.
The following programming and
configuration functions are covered:
1. Measurement type: pH Select pH,
ORP, Redox, Ammonia, Fluoride,
Custom ISE
2. Preamp location: Analyzer Identify
preamp location
3. Filter: 4 sec Override the default
input filter, enter 0-999 seconds
4. Reference Z: Low Select low or high
reference impedance
5. Sensor wiring scheme: Normal or Reference to Ground
To configure the ORP measurement board, access the Program screen by pressing
ENTER/MENU from the main screen and then select the Measurement tab.
1. To Select a Measurement type, select ORP and press ENTER/MENU.
2. To program the Preamp
location, select Analyzer or
Sensor/JBox, and press ENTER/MENU.
3. To Override the default input
44
Programming Measurements
Section 6: Programming Measurements
Instruction Manual
April 2017
LIQ-MAN-56
Содержание Rosemount 56
Страница 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Страница 2: ......
Страница 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Страница 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Страница 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Страница 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Страница 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Страница 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Страница 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...