Your instrument purchase from Emerson is one of the finest available for your particular application. These instruments
have been designed, and tested to meet many national and international standards. Experience indicates that its perform-
ance is directly related to the quality of the installation and knowledge of the user in operating and maintaining the instru-
ment. To ensure their continued operation to the design specifications, personnel should read this manual thoroughly
before proceeding with installation, commissioning, operation, and maintenance of this instrument. If this equipment is
used in a manner not specified by the manufacturer, the protection provided by it against hazards august be impaired.
• Failure to follow the proper instructions august cause any one of the following situations to occur: Loss of life; per-
sonal injury; property damage; damage to this instrument; and warranty invalidation.
• Ensure that you have received the correct model and options from your purchase order. Verify that this manual cov-
ers your model and options. If not, call 1-800-854-8257 or 949-757-8500 to request correct manual.
• For clarification of instructions, contact your Rosemount representative.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Use only qualified personnel to install, operate, update, program and maintain the product.
• Educate your personnel in the proper installation, operation, and maintenance of the product.
• Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes.
Only connect the product to electrical sources specified in this manual.
• Use only factory documented components for repair. Tampering or unauthorized substitution of parts and proce-
dures can affect the performance and cause unsafe operation of your process.
• All instrument enclosures must be closed and protective covers must be in place unless qualified personnel are per-
forming maintenance.
WARNING
RISK OF ELECTRICAL SHOCK
Equipment protected throughout by double insulation.
•Installation and servicing of this product august expose personnel to dangerous voltages.
•Main power wired to separate power source must be disconnected before servicing.
•Do not operate or energize instrument with case open!
•Signal wiring connected in this box must be rated at least 240 V for European mains operation.
•Non-metallic cable strain reliefs do not provide grounding between conduit connections! Use grounding type bush-
ings and jumper wires.
•Unused cable conduit entries must be securely sealed by non-flammable closures to provide enclosure integrity in com-
pliance with personal safety and environmental protection requirements. Unused conduit openings must be sealed with
Type 4X or IP66 conduit plugs to maintain the ingress protection rating (Type 4X)
•Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other
applicable national or local codes.
•Operate only with front panel fastened and in place.
•Safety and performance require that this instrument be connected and properly grounded through a three-wire power
source.
•Proper use and configuration is the responsibility of the user.
Essential Instructions
Read this page before proceeding
Содержание Rosemount 56
Страница 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Страница 2: ......
Страница 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Страница 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Страница 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Страница 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Страница 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Страница 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Страница 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...