Section 9: Alarm Relay Functions
Instruction Manual
April 2017
LIQ-MAN-56
68
Alarm Relay Functions
9.2.2
Setup
Access high/low concentration (setpoint) alarms by pressing ENTER/MENU from the main
screen and then Program/Relays/Configure Relay. From the main relay programming screen,
program this feature as follows:
1.
Relay:
Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2.
Type:
Select Setpoint as the relay type and press ENTER/MENU.
3.
Assign:
Assign S1 (sensor 1), S2 (sensor 2 if available) or other available parameters to
the designated relay and press ENTER/MENU.
4.
Logic:
Set High for a high setpoint or Low for a low setpoint and press ENTER/MENU.
5.
Setpoint:
Enter the desired setpoints value. Press ENTER/MENU.
6.
Deadband:
To set deadband as a measurement value, enter the change in the process
value needed after the relay deactivates to return to normal (and thereby preventing re-
peated alarm activation). Press ENTER/MENU.
7. Select
NEXT.
Press ENTER/MENU to advance to the next setup screen.
8.
Normal state:
Set the normal alarm condition as Open or Closed and press
ENTER/MENU. Program the normal state to define the desired alarm default state to nor-
mally open or normally closed upon power up.
* See Appendix 1 – HART and Device Variables
9.3
Delay Timer
9.3.1
Description
Delay Timer is a concentration control scheme which delays live measurement after chemical
addition using (one or all four of) of the 56 alarm relays. This ensures sufficient mixing time
in a vessel or recirculation loop before live sensor measurement, preventing unmixed
readings that might cause overshooting. Relay On time and Delay times are field-
programmable. Typical applications that would utilize the Delay Timer are: concentration
control in vessels, pH adjustments for neutralization and endpoint control for oxidation-
reduction reactions.
Relay Function
Limits and Selections
Default
Setpoint
NA
NA
Logic
Low/High
High
Setpoint
*
*
Deadband
*
0.000
On time
0 to 999.9 min
0 min
Delay time
0 to 999.9 min
0 min
Normal state
Close/Open
Open
Table 9-2. Defaults and programmable limits
Содержание Rosemount 56
Страница 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Страница 2: ......
Страница 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Страница 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Страница 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Страница 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Страница 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Страница 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Страница 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...