Instruction Manual
Section 9: Alarm Relay Functions
LIQ-MAN-56
April 2017
Alarm Relay Functions
73
Interval time has expired, the analyzer activates hold mode on the assigned measurement
and the relay is energized for the On time period.
A schematic of the Interval timer operation is shown:
9.6.2
Setup
Access Interval timer by pressing ENTER/MENU from the main screen and then
Program/Relays/ Configure Relay. From the main relay programming screen, program this
feature as follows:
1.
Relay:
Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2.
Type:
Select Interval timer as the relay type and press ENTER/MENU.
3.
Interval time:
Enter the time in hours (XX.X hours) between complete interval cycles.
4.
On time:
Enter the time in minutes (X.X min) for the relay to remain energized. The
assigned measurement value will be on hold during this time.
5.
Recovery time:
Enter the time in minutes (XX min) before the process is restored and
live measurements can resume.
6.
*Hold while active:
Select which sensors outputs should be on hold (S1, S2 or both)
during the interval timer activation time. Press ENTER/MENU.
7. Select
NEXT
Press ENTER/MENU to advance to the next setup screen.
8.
Normal state:
Set the normal alarm condition as Open or Closed and press
ENTER/MENU. Program the normal state to define the desired alarm default state to
normally open or normally closed upon power up.
*56 units with software ver. 2.1X and greater allow override of interval timer to ensure that all relays and outputs are held if desired.
Figure 9-5. interval timer Alarm operation
time
interval time
on time
hold
recovery
time
Relay Function
Limits and Selections
Default
Interval timer
NA
NA
Interval time
0 to 999.9 hr
24.0 hr
On time
0 to 999.9 sec
10 sec
Recovery time
0 to 999 sec
60 sec
Hold while active
0 to 999 sec
0 sec
Normal state
Sensor 1, Sensor 1, both
Sensor 1
Table 9-6. Defaults and programmable limits
Содержание Rosemount 56
Страница 1: ...Instruction Manual LIQ MAN 56 Rev D April 2017 Rosemount 56 Advanced Dual Input Analyzer ...
Страница 2: ......
Страница 10: ...iv Table of Contents Instruction Manual April 2017 LIQ MAN 56 ...
Страница 30: ...Section 2 Installation Instruction Manual April 2017 LIQ MAN 56 20 Installation ...
Страница 40: ...30 Wiring Section 3 Wiring Instruction Manual April 2017 LIQ MAN 56 ...
Страница 64: ...Section 6 Programming Measurements Instruction Manual April 2017 LIQ MAN 56 54 ...
Страница 76: ...Section 8 Time Proportional Control Instruction Manual April 2017 LIQ MAN 56 66 ...
Страница 124: ...Section 13 Maintenance Instruction Manual April 2017 LIQ MAN 56 114 ...
Страница 126: ...Section 14 Return of Material Instruction Manual April 2017 LIQ MAN 56 116 ...